Intelligent Motion Systems UDAS-1001E User Manual

Page 13

Advertising
background image

UDAS-1001E Series Data Acquisition Systems

1-6

If an analog input is used for the trigger, specify the trigger level, analog input channel, and

slope. Use analog output channel 1 for setting the trigger level.

When an input signal on the selected channel of the proper slope (rising or falling voltage) crosses

the trigger level, the conversion process begins.

Stopping Conversions
To stop a conversion, use either a software command, an N samples counter, or a combination of a

trigger and an N samples counter.

Use the N samples counter to specify the number of samples that you want to capture.

When the trigger is used to stop a conversion, and the N samples counter is used, a post-

triggering function occurs. In this mode, software is used to start conversions and continue

indefinitely until a trigger event occurs. When a trigger event occurs, N additional samples are

captured.

Rate Generator

Rate Generator

The UDAS unit contains an on-board 24-bit rate generator for analog input pacing. The base clock

for the rate generator is an 8 MHz clock derived from a crystal which provides 125 ns resolution

and a frequency range of 0.4768371 Hz to 4 MHz.

The I/O connector has a rate generator output pin that is driven by either the Analog Input Rate

Generator or the Analog Output Rate Generator.

Channel Scanner

Channel Scanner

When the channel scanner is enabled, immediately after the conversion on the current channel

finishes, the next analog input channel to be converted is automatically selected.

The UDAS device incorporates a 0 to N scanner. This allows analog conversions on all of the input

channels from channel 0 to channel N, where the value of N is programmable.

The scanner selects each channel for conversion, beginning with channel 0, incrementing to

channel N. After the conversion on channel N is complete, the scanner selects channel 0 and the

process repeats.

Note:

When using analog input channel 1 for a trigger, do not use this channel

for any other purpose.

Advertising