Intelligent Motion Systems UDAS-1001E User Manual

Page 62

Advertising
background image

C-7

Analog Output Channel 0 Offset and Gain

Analog Output Channel 0 Offset and Gain

1.

Verify the termination panel is attached to a 50-pin connector of a UDAS unit. Begin with step

2 if the unit has built-in termination.

2.

Attach a multimeter to analog output channel 0.

3.

Use SYSCHECK or other software to set analog output channel 0 to -10.000 V.

4.

Adjust Pot R26 for -10.000 V (±0.2 mV).

5.

Set analog output channel 0 to 9.9951 V.

6.

Adjust Pot R36 for 9.9951 V (±0.2 mV).

7.

Repeat these steps to verify the results.

Analog Output Channel 1 Offset and Gain

Analog Output Channel 1 Offset and Gain

1.

Verify the termination panel is attached to a 50-pin connector of a UDAS unit.

2.

Attach a multimeter to analog output channel 1.

3.

Use SYSCHECK or your calibration software to set analog output channel 1 to -10.000 V.

4.

Adjust Pot R24 for -10.000 V (±0.2 mV).

5.

Set analog output channel 1 to 9.9951 V.

6.

Adjust Pot R25 for 9.9951 V (±0.2 mV).

7.

Repeat these steps to verify the results.

Advertising