Analog Input and Output Calibration Procedures

C-4

.

7.

Set the calibrated input to 9.5068 V.

8.

Adjust Pot R46 for 9.5068 V.

9.

Repeat the procedure and verify the adjustments are correct.

10.

Set the precision voltage source to -9.95117 V and check for a reading of -9.95117.

Adjust Offset and Gain for ± 5Volt Input Range

Adjust Offset and Gain for ± 5Volt Input Range

1.

Verify the termination panel is connected to a 50-pin connector of the UDAS unit. Begin with

step 2 if the unit has built-in termination.

2.

Attach a voltage calibrator to channel 0 with all other channels grounded.

3.

Run a program, such as SYSCHECK, to configure the UDAS for a range of ±5 V and gain of 1.

4.

Set calibrated input to 0 V, or short channel 0 to analog ground.

5.

Set SYSCHECK or other program to average 50 readings. Manually or automatically take

readings repeatedly.

6.

Adjust Pot R39 until the software displays 0 V.

7.

Set the calibrated input to 4.5734 V.

8.

Adjust Pot R40 for 4.5734 V.

Note:

On all adjustments, there is likely a range of adjustment resulting in the proper

reading. Determine this range and leave the adjustment in the center of the range.

Note:

These adjustment steps are interactive and may require several iterations.

Note:

If the value you read is more than 0.039 V different from -9.95117, then the

following Alternate Bipolar Calibration must be used.

Note:

On all adjustments there will be a range of adjustment that results in the proper

reading. Determine this range and leave the adjustment in the center of the range.