❑ Inspect and clean the stability lock gears. Replace if worn.
❑ Clean upholstery and armrests.
❑ Inspect charger AC power cord for damage. Replace if necessary.
❑ Check that wiring is routed and secured properly to ensure that wiring does NOT 

become entangled and damaged.

❑ Inspect electrical components for signs of corrosion. Replace if corroded or damaged.
❑ Check that all labels are present and legible. Replace if necessary.
❑ Ensure casters are free of debris.

Troubleshooting - Mechanical

Troubleshooting - Electrical

NOTE: For additional troubleshooting information and explanation of error codes, refer to the 
individual Electronics Manual supplied with each wheelchair

SPJ+, SPJ+ w/PSS or SPJ+ w/ACC Joysticks

The joystick information gauge and the service indicator give indications of the type of 
fault or error detected by the control module. When a fault is detected, the wheelchair 
may stop and not drive. The LEDs on the information gauge may flash in a particular 
pattern or the service indicator light will flash. The number or type of flashes indicates the 
nature of the error. If multiple errors are found, only the first error encountered by the 
control module will be displayed.

WHEELCHAIR VEERS

LEFT/RIGHT

SLUGGISH

TU

RN/

PERFORMANCE

CA

STERS

FLUTTER

SQ

UEAKS AND RA

TTLES

LOOSENESS

IN WHEELCHAIR

WHEELCHAIR 3

WHEELS

SOLUTIONS

X

X

X

If pneumatic, check tires for correct and equal pressure.

X

X

X

X

Check for loose stem nuts/bolts.

X

X

Check that casters contact ground at the same time.