SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING

Part No 1143153

29

Pronto® M91™Base with SureStep®

Information Gauge Display Diagnostics

Service Indicator Light Diagnostics

DISPLAY

DESCRIPTION

DEFINITION

COMMENTS

All LEDs are off.

Power is off.

All LEDs are on.

Power is on.

Fewer than three LEDs on
implies reduced battery
charge.

Left RED LED is flashing.

Battery charge is low.

The batteries should be
charged as soon as
possible.

Left to Right “chase”
alternating with steady
display.

Joystick is in
programming, inhibit and/
or charging mode.

The steady LEDs indicate
the current state of the
battery charge.

All LEDs are flashing slowly. Joystick has detected Out-

of-Neutral-at-Power-Up
mode.

Release the joystick back
to Neutral.

NUMBER

OF

FLASHES

ERROR CODE

DESCRIPTION

POSSIBLE SOLUTION

1

User Fault

Release joystick to neutral and try again.

2

Battery Fault

Charge the batteries. Refer to Charging Batteries on page 63.
Check that battery cables are connected properly. Refer to Connecting/
Disconnecting the Battery Wiring Harness on page 60.
If necessary, replace batteries. Refer to Installing/Removing the Batteries
on page 57.

3

Left Motor Fault

Contact Invacare/Dealer for service.

4

Right Motor Fault

Contact Invacare/Dealer for service.

5

Left Park Brake Fault

Contact Invacare/Dealer for service.

6

Right Park Brake Fault Contact Invacare/Dealer for service.

7

Remote Fault

Check to make sure joystick is connected properly.
Contact Invacare/Dealer for service.

8

Controller Fault

Contact Invacare/Dealer for service.

9

Communications Fault Contact Invacare/Dealer for service.

10

General Fault

Contact Invacare/Dealer for service.

11

Incompatible or

incorrect Remote

Wrong type of remote connected. Contact Invacare/Dealer for service.

Information

Gauge

Display