Model 5100-XX-IT Toxic Gas Sensor Module

Page: 12

4.5 MODULE ADDRESS SWITCH

For digital interface applications the module address switch (or Modbus node) Figure 4-2 must be set per
Table 4-3:

P

OSITION

A

DDRESS

P

OSITION

A

DDRESS

1 Sensor

1 9 Sensor

09

2

Sensor 2

A

Sensor 10

3

Sensor 3

B

Sensor 11

4

Sensor 4

C

Sensor 12

5

Sensor 5

D

Sensor 13

6

Sensor 6

E

Sensor 14

7

Sensor 7

F

Sensor 15

8 Sensor

8 0

Software

Menu

Table 4 - 3: Sensor Module Address

Switch Positions

NOTES

For Sentry applications only sensor addresses 1-8 are allowed. If using Modbus output sensor

addresses 1-15 are available. Position 0 allows the Modbus Address to be set by software

menu, in the range 16-254.


Figure 4 - 2: Transmitter Face Plate