Model 5100-XX-IT Toxic Gas Sensor Module

Page: 42

Command 138: Apply ZERO Gas

This command will signal the module to accept the present sensor output to as the ZERO gas condition.

Request

Data

Bytes:

None

Command Specific Response Data Bytes: None


Command 139: Apply SPAN Gas

This command will signal the module to accept the present sensor output to as the SPAN gas condition.

Request

Data

Bytes:

None

Command Specific Response Data Bytes: None


Command 141: Force Gas Value

Request Data Bytes:

Byte

Format

Description

0

Unsigned-8

Sets the Force Gas Value, range 0 to 100

Command Specific Response Data Bytes:

Byte

Format

Description

0

Unsigned-8

Returns new Force Gas Value


Command 142: Reset Force Gas Value

This command will signal the module to set the Force Gas Value to 0.

Request

Data

Bytes:

None

Command Specific Response Data Bytes: None


Command 143: Read MODBUS RTU Register

This command allows the reading of sensor module data as defined in the MODBUS RTU register map.

Request Data Bytes:

Byte

Format

Description

0-1

Unsigned-16

Address of Requested Modbus Register
40001 = lowest address
65535 = highest address



Command Specific Response Data Bytes:

Byte

Format

Description

0-1

Unsigned-16

Returns contents of selected Modbus Register


Command 144: Set Transducer Serial Number

This command allows user to add a device specific serial number, if required. This is different from the serial
number assigned by the Manufacturer.

Request Data Bytes:

Byte

Format

Description

0-2

Unsigned-24 Serial

Number

Command Specific Response Data Bytes:

Byte

Format

Description

0-2

Unsigned-16

Returns new Serial Number