Model 5100-XX-IT Toxic Gas Sensor Module

Page: 43

Command 145: Set Calibration Mode output current

Request Data Bytes:

Byte

Format

Description

0-1

Unsigned-16

Current output required during calibration, in
micro-amperes

Command Specific Response Data Bytes:

Byte

Format

Description

0-1

Unsigned-16

Returns new Calibration current value


Command 146: Set Trouble Mode output current

Request Data Bytes:

Byte

Format

Description

0-1

Unsigned-16

Current output required to indicate Trouble, in
micro-amperes

Command Specific Response Data Bytes:

Byte

Format

Description

0-1

Unsigned-16

Returns new Trouble current value


Command 149: Set Gas Factor

Request Data Bytes:

Byte

Format

Description

0

Unsigned-8

Sets the Gas Factor value, range 30-200

Command Specific Response Data Bytes:

Byte

Format

Description

0

Unsigned-8

Returns new Gas Factor value

Command 150: Write MODBUS RTU Register

This command allows the writing of sensor module data as defined in the MODBUS RTU register map.
Request Data Bytes:

Byte

Format

Description

0-1

Unsigned-16

Address of Designated Modbus Register
40001 = lowest address
65535 = highest address

2-3

Unsigned-16

Data for Designated Modbus Register

Command Specific Response Data Bytes:

Byte

Format

Description

0-1

Unsigned-16

Returns MODBUS register value

Version 1.20 Updated 30 January 2012