S.P. Function - RS-485 Baud Rate Adjust

Table 5-3: Set-Up Configuration (Continued)

Enter

*38400

Press [E] to select or [] or [] to select another

Enter

Address

S.P. Function - RS-485 Address Adjust

Enter

ACK

New address selected

Enter

Addr:016

Use  or  keys to enter new address

RS-485 Adjustment Example

Enter

Address

S.P. Function - RS-485 Address Adjust

*E

Enter

*Track

* = Current selection

Enter

Track

Use  or  keys to select Track, C2.50mA, C4.00mA

Track = Output during calibration tracks the calibration gas,
C2.50mA = Output during calibration is 2.50 mA, C4.00mA =
Output during calibration is 4.0 mA

Down

CalibOut

S.P. Function - Output during Calibration Adjust

Mode

Calib

S.P. Function - Calib Adjust

Enter

ACK

Momentary Acknowledge of new Set Point

Enter

4mA

Selects 4 mA

Enter

Out: 4mA

Use  or  keys to select 4 mA or 20mA

4-20 mA Adjustment Example

Enter

Calib

S.P. Function - Calib Adjust

*D

*100 PPM

Select [E] to select or  or  to select another and
press [E]

Enter

100 PPM

Select [E] to select or  or  to select another and
press [E]

Down

USER

Press [E] if selecting User adjustable range

Gas Range Adjustment Example

Enter

Range

S.P. Function - Range Adjust

*C

Down

10 PPM

Press [E] if selecting 0-10 PPM range

Enter