Sierra Monitor Corporation 5100-06-IT User Manual

Page 45

Advertising
background image

Model 5100-XX-IT Toxic Gas Sensor Module

Page: 41

Command 132: Set Warning Level

Request Data Bytes:

Byte

Format

Description

0

Unsigned-8

Warning level, range 0 to 60. Default is 20.

Command Specific Response Data Bytes:

Byte

Format

Description

0

Unsigned-8

Returns new Warning Level.


Command 133: Set Alarm Relay Action

Request Data Bytes:

Byte

Format

Description

0

Unsigned-8

0=Latching (default)
2=Non-Latching

Command Specific Response Data Bytes:

Byte

Format

Description

0

Unsigned-8

Returns new Alarm Relay Action value


Command 134: Set Warning Relay Action

Request Data Bytes:

Byte

Format

Description

0

Unsigned-8 0=Latching

2=Non-Latching (default)

Command Specific Response Data Bytes:

Byte

Format

Description

0

Unsigned-8

Returns new Warning Relay Action value


Command 135: Reset Alarms

This command will reset any alarm relays that have been latched, providing the alarm condition is no longer
present on the sensor.

Request Data Bytes:

None

Command Specific Response Data Bytes: None


Command 136: Abort Calibration

This command aborts the calibration procedure.
Request

Data

Bytes:

None

Command Specific Response Data Bytes: None


Command 137: Set Calibration Gas Level

Request Data Bytes:

Byte

Format

Description

0

Unsigned-8

Calibration Gas Level, range 0 to 100

Command Specific Response Data Bytes:

Byte

Format

Description

0

Unsigned-8

Returns new Calibration Gas Level




Advertising
This manual is related to the following products: