Site Wiring Requirements,

Continued

DataNET Channels

The DataNET channel connections can be either copper wire or fiber
optic cable. The wiring method and cable type for DataNET is dependent
upon the site requirements. For cable requirements see Table 1-1.

The following Table should be used when evaluating or selecting cable
for an existing or new DataNET installation.

Table 1-1:
Recommended DataNET
Cable

Cable Type

Vendors & P/N

Installation Notes

Fiber Optic

MM 62.5/125; 2 Fiber
Conductors

SEA P/N 1270002-002

Industry Available from
Multiple Vendors

New Installation as
an alternative to
copper cable

Max distance from
Analyzer to ANG or
between Hubs
1520 meters (5000
ft)

2- Conductor Copper
Wire

SEA P/N 1686002-001

Belden P/N 9182

For

Belden Armored Cable

order

SEA P/N 1680000-002

Recommended for
Single Channel

Max distance from
Analyzer to ANG or
between Hubs
1520 meters (5000
ft)

4-Conductor Copper
Wire

SEA P/N 1681000-003

Belden P/N SSD1743

Recommended for
Dual Channel

Max distance from
Analyzer to ANG or
between Hubs
1520 meters (5000
ft)

8

Installation

2000592-001