Kodak 6B6640 User Manual

Page 25

Advertising
background image

Troubleshooting

March 2001

2-9

1003

Unable to load default
configuration file:_filename_.

The default configuration file was
deleted or renamed or moved.

Restore the file manually or
reinstall the software.

1004

Unable to set values
in new configuration file.

An internal error occurred with the
software.

Restart the application.

1005

Unable to save configuration
file:_filename_.

The system is trying to save a file
but there is not enough disk
space.

Remove any unnecessary files to
increase free disk space.

1006

Configuration file field:_field_
is invalid.

The user entered an incorrect
value in a configuration file.

• Enter the correct value.
• Reinstall the application.

1007

Unable to remove
directory:_directory_.

The user does not have the
authority to update the directory.

Verify that the user has update
authority on the directory.

1101

Density DMIN/DMAX points
are invalid.

Density points are invalid.

Reread density values on
densitometer or make a new
calibration print.

1102

Density data is not
monotonic increasing.

Density data is invalid.

Reread density values on
densitometer or make a new
calibration print.

1103

Exceeded maximum number
of allowable data errors.

Density data is invalid.

Reread density values on
densitometer or make a new
calibration print.

1104

Asymmetric Density
Groupings are invalid.

Density data is invalid.

Reread density values on
densitometer or make a new
calibration print.

1201

Unknown response from
densitometer.

During initialization, the
application received an unknown
response from the densitometer.

Verify that the baud rate of the
densitometer matches that of the
applications configuration.

Code

Status Message

Possible Cause/Subsystem

Possible Solution