Chapter 3

Signal Connections

© National Instruments Corporation

3-9

DAQCard-1200 User Manual

Figure 3-3. Differential Input Connections for Grounded Signal Sources

With this type of connection, the instrumentation amplifier rejects both the
common-mode noise in the signal and the ground-potential difference
between the signal source and the DAQCard-1200 ground, which is shown
as V

cm

in Figure 3-3.

Differential Connections for Floating Signal Sources

Figure 3-4 shows how to connect a floating signal source to a
DAQCard-1200 that is configured for DIFF input. Configuration
instructions are included in the

Analog Input Configuration

section of

Chapter 2,

Installation and Configuration

.

+

+

Grounded

Signal

Source

Vm

Measured

Voltage

Vs

+

Front I/O Connector

DAQCard-1200 in DIFF Configuration

1

3

5

7

2

4

6

8

ACH 0

AGND

11

ACH 2

ACH 4

ACH 6

ACH 1

ACH 3

ACH 5

ACH 7

Vcm

+

Common-Mode
Noise,
Ground
Potential