Appendix B

Differences among the Lab-PC+, the DAQPad-1200, and the DAQCard-1200

DAQCard-1200 User Manual

B-2

© National Instruments Corporation

The following table summarizes the differences among the three products.

Functionality

Lab-PC+

DAQPad-1200

DAQCard-1200

Bus resource allocation

Jumpers

Software

Software (automatic)

Calibration

Potentiometers

Software

Software

DMA

Uses DMA

Cannot use DMA

Cannot use DMA

Waveform generation

Can do waveform
generation on one
DAC and
immediate updates
on the other DAC

Can do waveform
generation on one
DAC and
immediate
updates on the
other

Should do either
immediate update on both
or waveform generation
on both but not a
combination. See
Chapter 4,

Theory of

Operation

, for more

information.

FIFO size

512 samples

1,024 samples

1,024 samples

Maximum single-channel
acquisition rate

83.3 K samples/s

100 K samples/s

100 K samples/s

Power-management modes

150 mA operating, 50 mA
power-down mode

Fuse

Regular 250 mA

Regular 250 mA

Thermal 1.0 A