Table 4-1. analog input settling time versus gain, Table 4-1, Analog input settling time versus gain -10 – National Instruments DAQCard-1200 User Manual
Page 63: Table 4-2, 1 sh, E 4-2

Chapter 4
Theory of Operation
4-10
© National Instruments Corporation
For multiple-channel data acquisition, observing the data acquisition rates
in Table 4-2 ensures 12-bit accuracy.
The recommended data acquisition rates in Table 4-2 assume that voltage
levels on all the channels included in the scan sequence are within range for
the given gain and are driven by low-impedance sources.
Note
If you use external conversions for single-channel acquisitions, the maximum
sampling rate is 90 kS/s.
Table 4-1. Analog Input Settling Time Versus Gain
Gain Setting
Settling Time Recommended
1, 2, 5, 10
10
µ
s max
20
12
µ
s typ, 15
µ
s max
50
25
µ
s typ, 30
µ
s max
100
60
µ
s typ, 80
µ
s max
Table 4-2. DAQCard-1200 Maximum Recommended Data Acquisition Rates
Acquisition Mode
Gain Setting
Sampling
Rate
Single channel
1, 2, 5, 10, 20, 50, 100
100.0 kS/s
Multiple channel
1, 2, 5, 10
100 kS/s
20
83.3 kS/s
50
40.0 kS/s
100