Theory of Operation

Chapter 3

SCXI-1163 User Manual

3-4

© National Instruments Corporation

Table 3-1. SCXIbus Equivalents for Rear Signal Connector

Rear Signal Connector

Signal

SCXIbus Equivalent

SERDATIN

MOSI

DAQD*/A

D*/A

SLOT0SEL*

INTR*

SERCLK

SPICLK

SERDATOUT

MISO

When the primary mode of operation is jumper configured to serial mode, the SCXI-1163
module converts the data acquisition board signals to open-collector signals on the backplane of
the SCXI chassis. For the data acquisition board to talk to a slot, the board must first assert the
SS* for that slot. To do this, assert INTR* low, write a 16-bit number over MOSI corresponding
to the desired slot (and chassis if you are using an SCXI-1001 chassis), and release INTR* high.
At this point, the SS* of the desired slot is asserted low and the data acquisition board can
communicate with the module in that slot according to the SPI protocol.

Interface and Routing Circuitry

Figure 3-3 shows a diagram of the SCXI-1163 and SCXIbus digital interface circuitry.

RESET*

MISO

SPICLK

INTR*

D*/A

MOSI

SS*

SCXIbus

SERDA

TIN

DAQD*/A

SLOT0SEL

SERCLK

Primary Mode

Control (Rear

Connector Routing)

SCXI Digital

Interface

Buffered Digital

Signal Controls

Buffered

Serial Data

Address

Handler

and

Registers

Rear Connector

Optically Isolated Inputs

SERDA

T

OUT

Parallel

Data

Output Mode

Control (Data

Routing)

Serial Data

Ouput Mode Select

Figure 3-3. Digital Interface and Routing Circuitry Block Diagram

Advertising