National Instruments 6023E User Manual

Page 35

Advertising
background image

Chapter 4

Signal Connections

© National Instruments Corporation

4-5

6023E/6024E/6025E User Manual

SCANCLK

DGND

Output

scan clock—this pin pulses once for each A/D conversion
in scanning mode when enabled. The low-to-high edge
indicates when the input signal can be removed from the
input or switched to another signal.

EXTSTROBE*

DGND

Output

External strobe—you can toggle this output under software
control to latch signals or trigger events on external devices.

PFI0/TRIG1

DGND

Input

Output

PFI0/Trigger 1—as an input, this is one of the
programmable function inputs (PFIs). PFI signals are
explained in the

Timing Connections

section in this chapter.

As an output, this is the TRIG1 (AI start trigger) signal.
In posttrigger data acquisition sequences, a low-to-high
transition indicates the initiation of the acquisition
sequence. In pretrigger applications, a low-to-high
transition indicates the initiation of the pretrigger
conversions.

PFI1/TRIG2

DGND

Input

Output

PFI1/Trigger 2—as an input, this is one of the PFIs.

As an output, this is the TRIG2 (AI stop trigger) signal. In
pretrigger applications, a low-to-high transition indicates
the initiation of the posttrigger conversions. TRIG2 is not
used in posttrigger applications.

PFI2/CONVERT*

DGND

Input

Output

PFI2/Convert—as an input, this is one of the PFIs.

As an output, this is the CONVERT* (AI convert) signal.
A high-to-low edge on CONVERT* indicates that an A/D
conversion is occurring.

PFI3/GPCTR1_SOURCE

DGND

Input

Output

PFI3/Counter 1 Source—as an input, this is one of the PFIs.

As an output, this is the GPCTR1_SOURCE signal. This
signal reflects the actual source connected to the
general-purpose counter 1.

PFI4/GPCTR1_GATE

DGND

Input

Output

PFI4/Counter 1 Gate—as an input, this is one of the PFIs.

As an output, this is the GPCTR1_GATE signal. This signal
reflects the actual gate signal connected to the
general-purpose counter 1.

GPCTR1_OUT

DGND

Output

Counter 1 Output—this output is from the general-purpose
counter 1 output.

Table 4-2. I/O Connector Signal Descriptions (Continued)

Signal Name

Reference

Direction

Description

Advertising