Signal Connections

© National Instruments Corporation

4-47

6023E/6024E/6025E User Manual

As an output, the GPCTR1_GATE signal monitors the actual gate signal
connected to general-purpose counter 1. This is true even if the gate is
externally generated by another PFI. This output is set to high impedance
at startup.

Figure 4-39 shows the timing requirements for the GPCTR1_GATE signal.

Figure 4-39. GPCTR1_GATE Signal Timing in Edge-Detection Mode

GPCTR1_OUT Signal

This signal is available only as an output on the GPCTR1_OUT pin.
The GPCTR1_OUT signal monitors the TC device general-purpose
counter 1. You have two software-selectable output options—pulse on TC
and toggle output polarity on TC. The output polarity is software-selectable
for both options. This output is set to high impedance at startup.
Figure 4-40 shows the timing requirements for the GPCTR1_OUT signal.

Figure 4-40. GPCTR1_OUT Signal Timing

GPCTR1_UP_DOWN Signal

This signal can be externally input on the DIO7 pin and is not available as
an output on the I/O connector. General-purpose counter 1 counts down
when this pin is at a logic low and counts up at a logic high. This input can
be disabled so that software can control the up-down functionality and

Rising-Edge

Polarity

Falling-Edge

Polarity

t

w

t

w

= 10 ns minimum

GPCTR1_SOURCE

GPCTR1_OUT

GPCTR1_OUT

(Toggle Output on TC)

(Pulse on TC)

TC