Sources -15, Figure 4-6 – National Instruments 6023E User Manual

Page 45

Advertising
background image

Chapter 4

Signal Connections

© National Instruments Corporation

4-15

6023E/6024E/6025E User Manual

Differential Connections for Nonreferenced or
Floating Signal Sources

Figure 4-6 shows how to connect a floating signal source to a channel
configured in DIFF input mode.

Figure 4-6. Differential Input Connections for Nonreferenced Signals

Figure 4-6 shows two bias resistors connected in parallel with the signal
leads of a floating signal source. If you do not use the resistors and the
source is truly floating, the source is not likely to remain within the
common-mode signal range of the PGIA. The PGIA then saturates, causing
erroneous readings.

+

+

Floating

Signal

Source

Input Multiplexers

V

m

Measured

Voltage

+

I/O Connector

AIGND

Bias

Current

Return

Paths

Bias
resistors
(see text)

ACH–

ACH+

AISENSE

Selected Channel in DIFF Configuration

PGIA

Programmable Gain

Instrumentation

Amplifier

Vs

Advertising
This manual is related to the following products: