Chapter 4

Signal Connections

6023E/6024E/6025E User Manual

4-24

ni.com

Power-up State

♦ 6025E only

The 6025E contains bias resistors that control the state of the digital I/O
lines PA<0..7>,PB<0..7>,PC<0..7> at power up. Each digital I/O line is
configured as an input, pulled high by a 100 k

Ω bias resistor.

You can change individual lines from pulled up to pulled down by adding
your own external resistors. This section describes the procedure.

Changing DIO Power-up State to Pulled Low

Each DIO line is pulled to V

cc

(approximately +5 VDC) with a 100 k

resistor. To pull a specific line low, connect between that line and ground
a pull-down resistor (R

L

) whose value gives you a maximum of 0.4 VDC.

The DIO lines provide a maximum of 2.5 mA at 3.7 V in the high state.
Using the largest possible resistor ensures that you do not use more current
than necessary to perform the pull-down task.

However, make sure the value of the resistor is not so large that leakage
current from the DIO line along with the current from the 100 k

Ω pull-up

resistor drives the voltage at the resistor above a TTL-low level of 0.4 VDC.
Figure 4-12 shows the DIO configuration for high DIO power-up state.

Figure 4-12. DIO Channel Configured for High DIO Power-up State with External Load

Example

A given DIO line is pulled high at power up. To pull it low on power up with
an external resistor, follow these steps:

1.

Install a load (R

L

). Remember that the smaller the resistance, the

greater the current consumption and the lower the voltage.

Device

Digital I/O Line

82C55

100 k

GND

R

L

+5 V