Chapter 3

Signal Connections

© National Instruments Corporation

3-9

DAQCard-700 User Manual

Differential signal connections reduce picked-up noise and increase common-mode signal and
noise rejection. With these connections, input signals can float within the common-mode limits
of the input instrumentation amplifier.

Differential Connections for Grounded Signal Sources

Figure 3-4 shows how to connect a ground-referenced signal source to a DAQCard-700 card
configured for DIFF input.

Ground-
Referenced
Signal
Source

Common-
Mode
Noise,
Ground
Potential,
and so on

Input Multiplexers

Instrumentation

Amplifier

V

m

Measured
Voltage

AIGND

-

+

-

+

V

s

V

cm

+

-

-

+

I/O Connector

DAQCard-700 Board in DIFF Configuration

3

5

7

17

4

6

8

18

1, 2

ACH<0..7>

ACH<8..15>

Figure 3-4. Differential Input Connections for Grounded Signal Sources

With this type of connection, the instrumentation amplifier rejects both the common-mode noise
in the signal and the ground potential difference between the signal source and the
DAQCard-700 ground (shown as V

cm

in Figure 3-4).