National Instruments SCXI-1127 User Manual

Page 11

Advertising
background image

© National Instruments Corporation

1-1

SCXI-1127/1128 User Manual

1

Installing and Configuring the
SCXI-1127/1128

The SCXI-1127 and SCXI-1228 modules are high-voltage
multiplexer/matrix modules. The SCXI-1127 is a high-density armature
relay module. The SCXI-1128 is a high-density solid-state relay module.
Both modules provide a 1

× 32, 2-wire multiplexer and several switch

matrix configurations on an SCXI platform. The SCXI-1127/1128 can also
operate as two 1

× 16 or four 1 × 8, 2-wire multiplexers (in independent

mode) allowing a single module to handle several multiplexing needs.

Through software-configurable switches, the module can also operate as a
1-wire 1

× 64 multiplexer, enabling you to make a large number of common

referenced measurements. It can also operate as a 1

× 64, 4-wire

multiplexer to make resistive measurements such as those needed for
thermistors and RTDs. With the use of special terminal blocks, the
SCXI-1127/1128 can become a 4

× 8 or 1 × 32, 2-wire matrix. A complete

list of possible configurations is given in Table 1-1.

To handle large channel counts, you can expand the size of the multiplexer
with additional SCXI-1127/1128 modules. These modules support the
analog bus through a special adapter, the high-voltage analog bus (HVAB)
backplane adapter in the SCXI-1357/1358 kit, at the rear of the module.
This adapter allows you to expand the HVAB without complicated wiring.
For example, you can connect two SCXI-1127/1128 modules using the
HVAB-backplane adapter to create a 1

× 64, 2-wire multiplexer.

Table 1-1. SCXI-1127/1128 Configurations

Scanning

Configuration

Multiplexer

Matrix

1-wire

1

× 64

N/A

2-wire

1

Ч 32

4

Ч 8 (also 2 Ч 16, 1 Ч 32)

4-wire

1

× 16

N/A

Advertising
This manual is related to the following products: