Figure 2-16, Temperature measurement wiring diagram -18 – National Instruments SCXI-1127 User Manual

Page 53

Advertising
background image

Chapter 2

Using the SCXI-1127/1128

SCXI-1127/1128 User Manual

2-18

ni.com

Figure 2-16. Temperature Measurement Wiring Diagram

The SCXI-1331 contains a cold-junction temperature sensor (CJS). This
sensor is a special channel on the SCXI-1127/1128 dedicated to measuring
the ambient temperature of the terminal block. This channel is always
scanned as a 2-wire channel. Including a CJS channel in the scan list is
optional. You can include a CJS channel at any position in the list with any
number of repetitions.

The SCXI-1331 temperature sensor outputs 0.2–0.024 V from 0–50 °C and
has an accuracy of ±0.5 °C over the 15–35 °C range and ±0.9 °C over the
0–15 °C and 35–50 °C ranges.

1

National Instruments software can convert a thermistor voltage to the
thermistor temperature. In LabVIEW, you can use the Convert Thermistor
Reading virtual instrument (VI) in the Data Acquisition»
Signal Conditioning
palette. If you are using Measurement Studio or

1

Includes the combined effects of the temperature sensor accuracy and the temperature difference between the temperature
sensor and any screw terminal. The temperature sensor accuracy includes tolerances in all component values, the effects
caused by temperature and loading, and self-heating.

SCXI-1127/1128

R

1

-

+

Iex

RTD

Thermistor

R

2

V

1

Thermocouple

Advertising