Chapter 2

Using the SCXI-1127/1128

SCXI-1127/1128 User Manual

2-10

ni.com

current and the other for sensing the voltage developed across the resistor,
as shown in Figure 2-9.

Figure 2-9. Signal Connections for a 4-Wire Resistance Measurement

Since ideally the impedance of the voltmeter is infinite, very little or no
current passes through the sense leads of the voltmeter. As a result, the
voltmeter reads the exact voltage developed across R

1

, and no parasitic

resistances are introduced into the measurement:

Therefore, a 4-wire configuration is a more accurate way of performing
resistance measurements than a 2-wire configuration.

Mixed Mode Configuration

The input mode of the SCXI-1127/1128 is configurable on a per channel
basis; therefore, you can connect a variety of signal types in 2-wire, 1-wire,
and 4-wire modes to the same module. An SCXI-1127/1128 can have
2-wire, 1-wire, and 4-wire channels connected to its front at the same time,
as shown in Figure 2-10.

V

SENSE

+

4-Wire

Measurement

R

P

V

R

1

R

1

I

EX

I

P

~0

I

P

~0

R

P

R

P

I

EX

R

P

V

SENSE

R

1measured

V

R

1

I

EX

--------

R

1

=

=