Figure 2-3. 2-wire wiring diagram, Figure 2-3, Wire wiring diagram -4 – National Instruments SCXI-1127 User Manual

Page 39

Advertising
background image

Chapter 2

Using the SCXI-1127/1128

SCXI-1127/1128 User Manual

2-4

ni.com

Figure 2-3 shows an example of a 2-wire configuration using the
SCXI-1331 terminal block. The diagram shows the 2-wire terminal block
connections for channels 2, 5, 24, and 27. Channel 2 measures a voltage
source (V

1

). The positive terminal of channel 2 is labeled +2 with the

negative channel labeled –2. The output of the multiplexer is available
through the OUT0 bus.

Figure 2-3. 2-Wire Wiring Diagram

You can connect different types of 2-wire inputs to the SCXI-1127. For
example, in Figure 2-3, a thermocouple (V

2

) is shown connected to

channel 27, and on channel 5 (R

1

) and 24 (R

2

) resistance measurements are

being made. Consult the

Making Temperature Measurements

section for

more information on temperature compensation measurements for making
accurate thermocouple measurements.

SCXI-1127

R

1

R

2

V

2

Thermocouple

Multiplexer Output

+

V

1

+

OUT0±