Slot and 12-slot multichassis configuration, Slot and 12-slot multichassis configuration -14 – National Instruments SCXI-1127 User Manual

Page 24

Advertising
background image

Chapter 1

Installing and Configuring the SCXI-1127/1128

SCXI-1127/1128 User Manual

1-14

ni.com

1.

Install the HVAB-backplane adapters and jumper blocks in the 12-slot
chassis as described in steps 1 through 3 in

Single 12-Slot Chassis

Configuration

.

2.

Connect the HV8-HV8 cable from the first chassis, normally the
connector behind slot 1, to the HVAB connector (behind slot 12) of the
next chassis.

3.

Connect the SH9MD-9MD cable from the AUX OUT connector of the
first chassis to the AUX IN connector (behind slot 5) of the next
chassis.

4.

Repeat steps 1 through 3 for each additional chassis.

Notes

Do not use the SH9MD-9MD cable behind slot 4 of the 2-slot HVAB-backplane

adapter on the 12-slot chassis.

An SCXI-1127/1128 is required in slot 4 of the 4-slot chassis and slot 12 of the 12-slot
chassis to establish communications with all the chassis in the system. If slot 4 is empty
in the 4-slot chassis or slot 12 is empty in the 12-slot chassis, the chassis will not operate.
It is this module that you must specify in MAX as the module cabled to the DMM.

If the 4-slot chassis is configured with less than four slots of the HVAB-backplane adapter,
use the right-most available HVAB connector to extend to the next chassis. In this instance,
right-most means the slot closest to slot 1 when the chassis is viewed from the back.

12-Slot and 12-Slot Multichassis Configuration

In addition to the items needed for a single-chassis system, you need the
following items to install this configuration:

SCXI-1358 multichassis expansion kit

HV8-HV8 cable

SH9MD-9MD cable

Three 8-position HVAB plugs

One 8-slot HVAB-backplane adapter

One 2-slot HVAB-backplane adapter

Two 1-slot HVAB-backplane adapters

Advertising