Chapter 1

Installing and Configuring the SCXI-1127/1128

© National Instruments Corporation

1-17

SCXI-1127/1128 User Manual

5.

Connect the HV8-BAN4 cable from the DMM to the HVAB connector
behind SCXI slot 4.

6.

Connect the SH9MD-9MD cable from the DMM to the AUX IN
connector.

7.

You can install any additional SCXI-1127/1128 modules in any slot
that has an HVAB-backplane adapter behind it.

DMM in PXI Slot 8 Configuration

To install your DMM in PXI slot 8, follow these steps:

1.

Install your DMM in PXI slot 8.

2.

In the SCXI subsystem, install the 2-slot HVAB-backplane adapter
behind SCXI slots 3 and 4.

3.

Install additional 1-slot HVAB-backplane adapters behind SCXI slots
1 and 2 if needed.

4.

Install the 8-position HVAB plugs to connect the HVAB-backplane
adapters as needed.

5.

Connect the HV8-BAN4 cable from the DMM to the HVAB connector
behind SCXI slot 4 of the SCXI subsystem.

6.

You can install any additional SCXI-1127/1128 modules in any slot
that has an HVAB-backplane adapter behind it.

Note

The SH9MD-9MD is not required for this configuration.

Connecting the SCXI-1127/1128 to the DMM in the PXI-1011 Chassis

You will need the following items to complete the installation:

PXI-1011 chassis with the SCXI-1127/1128 modules installed

HV8-BAN4 cable

DMM

1/8 in. flathead screwdriver

Consult your PXI-1011 and DMM documentation for additional
instructions and warnings. Your SCXI-1127/1128 modules and any other
SCXI modules should already be installed in the chassis according to their
installation instructions. Before you proceed with your connections, make
sure you turn off y our SCXI chassis.