Figure 2-5. Differential Input Connections for Ground-Referenced Signals

With this connection type, the instrumentation amplifier rejects both the
common-mode noise in the signal and the ground potential difference
between the signal source and the NI PXI-7831R ground, shown as V

cm

in Figure 2-5. In addition, the instrumentation amplifier can reject
common-mode noise pickup in the leads connecting the signal sources to
the device. The instrumentation amplifier can reject common-mode signals
as long as V+

in

and V–

in

(input signals) are both within their specified input

ranges. Refer to Appendix A,

Specifications

, for more information about

input ranges.

Differential Connections for Nonreferenced or
Floating Signal Sources

Figure 2-6 shows how to connect a floating signal source to a channel on
the NI PXI-7831R configured in DIFF input mode.

DIFF Input Mode Selected

+

+

+

+

V

cm

V

s

Ground-

Referenced

Signal

Source

Common-

Mode

Noise and

Ground

Potential

I/O Connector

AISENSE

AIGND

V

m

Measured

Voltage

Instrumentation

Amplifier

x8 Channels

AI+

AI–