Chapter 4

Connecting Signals

© National Instruments Corporation

4-9

NI 6013/6014 User Manual

setting of the amplifier. The amplifier output voltage is referenced to the
device ground. The device ADC measures this output voltage when it
performs A/D conversions.

Connecting Analog Input Signals

The following sections discuss the use of single-ended and differential
measurements and make recommendations for measuring both floating
and ground-referenced signal sources.

Figure 4-3 summarizes the recommended input configuration for both
types of signal sources.

Figure 4-3. Summary of AI Connections

Input

Input

Floating Signal Source

(Not Connected to Building Ground)

Grounded Signal Source

Differential

(DIFF)

Single-Ended —

Nonreferenced

(NRSE)

Examples
• Ungrounded Thermocouples
• Signal Conditioning with

Isolated Outputs

• Battery Devices

Example
• Plug-in Instruments with Nonisolated

Outputs

+

+

V

1

ACH

AISENSE

AIGND

+

+

V

1

ACH(+)

ACH(–)

AIGND

+

+

V

1

ACH(+)

ACH(–)

AIGND

R

See text for information on bias resistors.

+

+

V

1

ACH

AISENSE

AIGND

R