Figure 4-32. gpctr1_source signal timing, Gpctr1_gate signal, Gpctr1_gate signal -37 – National Instruments NI 6014 User Manual

Page 63

Advertising
background image

Chapter 4

Connecting Signals

© National Instruments Corporation

4-37

NI 6013/6014 User Manual

Figure 4-32 shows the timing requirements for GPCTR1_SOURCE.

Figure 4-32. GPCTR1_SOURCE Signal Timing

The maximum allowed frequency is 20 MHz, with a minimum pulse width
of 23 ns high or low. There is no minimum frequency limitation.

The 20 MHz or 100 kHz timebase normally generates GPCTR1_SOURCE
unless you select some external source.

GPCTR1_GATE Signal

Any PFI pin can externally input the GPCTR1_GATE signal, which
is available as an output on the PFI4/GPCTR1_GATE pin.

As an input, GPCTR1_GATE is configured in edge-detection mode.
You can select any PFI pin as the source for GPCTR1_GATE and configure
the polarity selection for either rising or falling edge. You can use the gate
signal in a variety of different applications to perform such actions as
starting and stopping the counter, generating interrupts, saving the counter
contents, and so on.

As an output, GPCTR1_GATE monitors the actual gate signal connected
to general-purpose counter 1, even if the gate is being externally generated
by another PFI. This output is set to high-impedance at startup.

t

w

= 23 ns minimum

t

p

= 50 ns minimum

t

w

t

w

t

p