Chapter 4

Connecting Signals

NI 6013/6014 User Manual

4-24

ni.com

ignores TRIG2 if it is asserted prior to the SC decrementing to zero. After
the selected edge of TRIG2 is received, the device acquires a fixed number
of scans and the acquisition stops. This mode acquires data both before and
after receiving TRIG2.

As an output, TRIG2 reflects the posttrigger in a pretriggered DAQ
sequence, even if the acquisition is being externally triggered by another
PFI. TRIG2 is not used in posttriggered data acquisition. The output is an
active high pulse with a pulse width of 50 to 100 ns. This output is set to
high-impedance at startup.

Figures 4-15 and 4-16 show the input and output timing requirements
for TRIG2.

Figure 4-15. TRIG2 Input Signal Timing

Figure 4-16. TRIG2 Output Signal Timing

Rising-Edge

Polarity

Falling-Edge

Polarity

t

w

= 10 ns minimum

t

w

t

w

= 50 to 100 ns

t

w