Figure 3-5. below-low-level analog triggering mode – National Instruments NI PCI-6110 User Manual

Page 25

Advertising
background image

Chapter 3

Hardware Overview

NI PCI-6110/6111 User Manual

3-6

ni.com

In below-low-level analog triggering mode, the trigger is generated when
the signal value is less than lowValue, as shown in Figure 3-5. HighValue
is unused.

Figure 3-5. Below-Low-Level Analog Triggering Mode

In above-high-level analog triggering mode, the trigger is generated when
the signal value is greater than highValue, as shown in Figure 3-6.
LowValue is unused.

Figure 3-6. Above-High-Level Analog Triggering Mode

In inside-region analog triggering mode, the trigger is generated when the
signal value is between the lowValue and the highValue, as shown in
Figure 3-7.