Chapter 4

Connecting Signals

© National Instruments Corporation

4-11

NI PCI-6110/6111 User Manual

Differential Connections for Ground-Referenced
Signal Sources

Figure 4-4 shows how to connect a ground-referenced signal source to a
channel on the NI PCI-6110/6111.

Figure 4-4. Differential Input Connections for Ground-Referenced Signals

With this type of connection, the PGIA rejects both the common-mode
noise in the signal and the ground potential difference between the signal
source and the device ground, shown as V

cm

in Figure 4-4.

PGIA

Ground-

Referenced

Signal

Source

Common-

Mode

Noise and

Ground

Potential

V

s

V

cm

+

+

I/O Connector

ACH0 Connections Shown

ACH0GND

ACH0–

ACH0+

+

+

V

m

Measured

Voltage

Instrumentation

Amplifier