7 objectives, Ix81 – Olympus IX81 User Manual

Page 31

Advertising
background image

26

IX81

Fig. 34

1

2

1

Oil-Proof Cap (UIS Series only)

(Fig. 34)

}By mounting the oil-proof cap (Type C1 or C2) on the tip of the applicable

objective, you can prevent penetration of immersion oil or water into the
objective. This will enable the objective to achieve its original performance,
so be sure to always mount the cap.

How to mount

Make sure which type, C1 or C2, should be used for the objective in use
by referring to the table below. Then fit the appropriate oil-proof cap @
into the mounting groove ² securely.

Applicable Objectives

Oil-proof
cap

Objective (Series 3)

Remark

Oil-proof
cap

Objective (Series 3)

Remark

C1

C2

UPlanFl100XO3

´´

100XO3Ph

´´

100XO3P

Provided with
microscope frame.
C1: 1 piece
C2: 2 pieces

C1

UApo20X3/340

´´

40X3/340

´´

40XOI3/340

´´

20XW3/340

´´

40XW3/340

UPlanApo10XW3

´´

Apo10XO3

´´

Apo60XW3

´´

Apo60XW3/IR

1 piece is provided
with each
objective.

UPlanFl60XOI3

´´

60XOI3Ph

´´

100XOI3

UPlanApo20XO3

´´

40XOI3

´´

40XOI3Ph

´´

100XOI3

´´

100XOI3Ph

PlanApo60XO3

´´

60XO3Ph

´´

100XO3

Caution when cleaning

When wiping off immersion oil or oil attached to the oil-proof cap, do not press the cap hard. Otherwise, the tip of the
objective will retract due to the buffer spring mechanism, causing the cap to come off and oil or water to penetrate into
the objective. Take care not to push the cap when wiping it.

4-7 Objectives

Advertising