Olympus IX81 User Manual

Page 40

Advertising
background image

35

Fig. 49

Fig. 50

Fig. 51

5

Cross-Nicol Adjustment

(Figs. 49 to 51)

1. Rotate the condenser’s turret for the BF (brightfield) light path (with no

optical element engaged in the light path).

2. Move the polarizer detaching lever 1 on the IX-LWPO polarizer to engage

the polarizer in the light path. (Fig. 49)

3. Engage the 10X objective in the light path, place an optimum specimen for

brightfield observation on the stage, bring the specimen into approximate
and remove the specimen out of the light path.

4. When the U-BI90CT, binocular observation tube is used, rotate the CT

turret 2 to position “CT” to engage the CT lens in the light path. (Fig. 50)
When the U-BI90/U-TBI90 binocular observation tube or U-TR30H-2
trinocular observation tube is used, remove an eyepiece and attach
the U-CT30 centering telescope.

5. Rotate the focusing ring 3 (or, with the centering telescope, rotate the

knurled section) to bring the objective’s exit pupil into focus. (Fig. 50)

6. Move the prism movement knob 7 of the DIC slider in the clockwise

direction around the axis until, the knob is stopped. A black interference
stripe then a rainbow-colored interference stripe will be observed. Here,
stop the knob at the position with which the black interference stripe can
be seen. (Figs. 46 & 51)

7. While observing the objective’s exit pupil, rotate the polarizer rotation/

clamping knob 4 on the polarizer unit horizontally until the black interfer-
ence stripe becomes darkest. This is the position for use of the polarizer.

(Figs. 49 & 51)

8. After determining the position, clamp the polarizer rotation/clamping knob

so that the polarizer will not rotate.

1

4

2

3

Advertising