Attaching the Camera on the Lower Back Port

}A C-mount video camera can be used by ordering the manufacturer to

install and adjust the IX2-LBPC C-mount for lower back port. Installing
the TV camera in this way allows you to use the space in front of the
microscope efficiently.
The size of the video camera is limited to 57 mm in the up or bottom
direction.

}Two kinds of back port prisms are available. Select either prism.
· IX2-PRLBP8: Back 80 % & Observation 20%, or Observation 100%
· IX2-PRLBP1: Back 100% or Observation 100%

4

}A C-mount video camera (used together with the U-CMT) or F-mount

video camera (used together with the U-FMT) can be used by ordering
the manufacturer to install and adjust the IX2-TVR TV port unit for bottom
panel.
Only a C-mount video camera can be used with the IX2-TVRAC motorized
TV port unit for bottom panel.
Even after modification, the binocular 100% observation or TV 100% ob-
servation are possible by setting the light path selector lever.

5

TV Observation

1. Adjusting the microscope frame

Turn on the light source and perform all the required adjustments for
observation.

2. Adjusting the video camera and monitor

Perform the required adjustments such as the white balance and color
adjustments.

# The center of the field of view of eyepiece may not match the center

of the monitor screen. This is due to the adjustment mechanism of
the video camera’s imaging plane and cannot be corrected.

3. Perfocality adjustment between the observed image and monitor image
· In accordance with the adjustment method of each TV adapter, adjust for

simultaneous focusing of the observed image and monitor image.

<<

TV Observation Magnification

>>

TV observation magnification

= Objective magnification x TV adapter lens magnification x TV monitor

size/Video camera’s image pickup device size