OTHER OBSERVATION METHODS

5-1 Phase Contrast Observation

}A phase contrast objective, phase contrast optical element and the U-CT30 centering telescope are required for phase

contrast observation (the U-BI90CT binocular observation tube .is not required).

}If a DIC slider, analyzer or polarizer is engaged in the light path, disengage it.

1

Phase Contrast Optical Elements and Applicable Objectives

With the IX2-LWUCDA2 and IX2-LWUCD Condenser

Optical Element

Indication

Applicable Objectives

IX-PHL (small)

PhL

UPlanFl4XPh

IX-PHC (small)

PhC

CPlan10XPh, LCAch20XPh, CPlanFl10XPh

IX-PH1 (small)

Ph1

UPlanFl10XPh, UPlanFl20XPh, LCPlanFl20XPh, UPlanApo10XPh

IX-PH2 (small)

Ph2

UPlanFl40XPh, LCPlanFl40XPh, LCPlanFl60XPh, UPlanApo20XPh, LCAch40XPh,
LUCPlanFl40XPh, SLCPlanFl40XPh

IX-PH3 (large)

Ph3

UPlanFl100XO3Ph, UPlanApo40XO

I3Ph, UPlanApo100XOI3Ph, PlanApo60XO3Ph

}Insert the optical element (small) in the 30 mm position and the optical element (large) in the 38 mm position.

For well observation, it is recommended to use the IX-PHC to obtain the phase contrast effect in a wide range of field of
view.

Optical Element

Indication

Applicable Objectives

IX-PHL (small)

PhL

UPlanFLN4XPh

IX-PHC (small)

PhC

CPlanN10XPh, LCAchN20XPh, CPlanFLN10XPh

IX-PH1 (small)

Ph1

UPlanFLN10X2Ph, UPlanFLN20XPh, LUCPlanFLN20XPh

IX-PH2 (small)

Ph2

UPlanFLN40XPh, LUCPlanFLN40XPh, LUCPlanFLN60XPh,
LCAchN40XPh, UPlanFLN60XOIPh

IX-PH3 (large)

Ph3

UPlanFLN100XO2Ph

UIS2 Series