}The following four methods are available for TV observation.
1. Attaching the video camera onto the side port
2. Attaching the video camera to the U-TR30H-2 trinocular observation tube H
3. Attaching the video camera onto the IX2-LBPC C-mount unit for lower back port (To be installed by the manufacturer)
4. Attaching the video camera onto the IX2-TVR/IX2-TVRAC bottom TV port unit (To be installed by the manufacturer)

1

Attaching the Camera on the Side Port

}The side port provides primary observation images that are brighter and higher in resolution than those obtained with the

trinocular observation tube H.

· Compatibility with various mounts from different manufacturers can be obtained by using the IX2-SPT straight photo tube

+ TV system for the BX series or by using a TV adapter for the BX series. However, this does not apply to the U-TVCAC and
U-DPT.

· For the installation of the IX2-SPT, see page 38.
· For the TV adapter systems for the BX series, refer to the instruction manual for the TV adapter to be used.

2

Attaching the Camera on the Trinocular Observation Tube H

(Fig. 59)

· Compatibility with various mounts from different manufacturers can be obtained by attaching the IX-SPT straight photo

tube on the U-TR30H-2 trinocular observation tube H and adding a TV adapter system for the BX series.

· A C-mount or F-mount video camera cam be attached by attaching the U-CMT C-mount adapter or U-FMT F-mount

adapter onto the IX-TVAD TV adapter.

Installation and Parfocal Adjustment of IX-TVAD TV Camera

1. Using the Allen screwdriver, loosen the clamping screw 1 on the trinocular

observation tube H.

2. Fit the IX-TVAD TV adapter 2 into the straight photo tube and tighten the

clamping screw 1. (Fig. 59)

3. Firmly screw the U-CMT C-mount adapter 3 or U-FMT F-mount adapter

4 into the IX-TVAD TV adapter 2. (Fig. 59)

4. Attach a compatible TV camera on the U-CMT or U-FMT.

<<

Parfocal Adjustment of Observation Image and Monitor Image

>>

1. Adjust the eyepiece to bring the specimen into focus.
2. Lightly loosen the clamping screws 1 and 5. (Fig. 59)
3. While observing the monitor image and holding the U-CMT 3 or U-FMT

4 stationary, rotate the IX-TVAD 2. Tighten the clamping screws 1 and
5 when the monitor image is in focus. (Fig. 59)

1

2

3

4

5