VXI VT1433B User Manual

Page 144

Advertising
background image

Dis-allowed Connections

Do NOT connect these pins on VT1432A/33B end of cabling. These signals and
supplies are provided for VXI Technology specified break out boxes and are
unspecified for other usage

Do not use:

15

I2C_EN

17

I2C_SDA

21

I2C_SCL

16

RFI GND/I2C Shield

18

RFI GND/I2C Shield

22

RFI GND/I2C Shield

19

BOB_EN

7

CAL_LOW

20

CAL_HIGH

23

-24 V Power

24

±24 V GND Return

25

+24 V Power

In general:

q

± DIFF n lines are the differential inputs for each channel. Shielded twisted-pair is
recommended.

q

RFI GND/Drain Shield n are the grounds for the shield on the twisted-pair for each
input channel. Connect at the VT1432A/33B end of the cable only.

q

RFI GND/Cable Shield are the grounds for a shield around the entire cable and the
ground points for making individual channels single-ended.

q

I2C_xxx supply control signals to the active break out boxes. Support is not provided
for any other use. These are not used with the VT3240/1A Voltage and Voltage/ ICP

®

break out boxes.

q

RFI GND/I2C Shield protects the analog input lines.

q

BOB_EN is another break out box control signal. Support is not provided for usage
with anything except certain VXI Technology-specified break out boxes.

q

CAL_HIGH/LOW are signal lines to send calibration signals to the
VXI Technology-specified break out boxes. The signals available on these lines are
not specified and their usage is discouraged.

q

±24V Power and GND supply power to the signal conditioning circuitry in the active
break out boxes and ICP

®

in the active ICP

®

break out box. The power available on

these lines is not specified and their usage is discouraged.

VT1433B User's Guide
Break Out Boxes

8-8