VXI VT1433B User Manual

Page 59

Advertising
background image

Table of Baseband Measurement Spans (part 4 of 6)
All values are in Hertz (Hz).

sample frequency

(Fs) —>

10000

102400

128000

number of

÷2 steps

with ÷5

without ÷5

with ÷5

without ÷5

with ÷5

without ÷5

0

7812.500000

39062.500000*

8000.000000

40000.000000*

10000.000000

50000.000000*

1

3906.250000

19531.250000

2000.000000

20000.000000

5000.000000

25000.000000

2

1953.125000

9765.625000

2000.000000

10000.000000

2500.000000

12500.000000

3

976.562500

4882.812500

1000.000000

5000.000000

1250.000000

6250.000000

4

488.281250

2441.406250

500.000000

2500.000000

625.000000

3120.500000

5

244.140625

1220.703125

250.000000

1250.000000

312.500000

1562.500000

6

122.070312

610.351562

125.000000

655.000000

156.250000

781.250000

7

61.035156

305.175781

60.250000

312.500000

78.125000

390.625000

8

30.517578

152.587891

30.125000

156.250000

39.062500

195.312500

9

15.258789

76.293945

15.625000

78.125000

19.531250

97.656250

10

7.629395

38.146973

7.812500

39.062500

9.765625

48.828125

11

3.814697

19.073486

3.906250

19.531250

4.882812

24.414062

12

1.907349

9.536743

1.953125

9.765625

2.441406

12.207031

13

0.953674

4.768372

0.976562

4.882812

1.220703

6.103516

14

0.476837

2.384186

0.488281

2.441406

0.610352

3.051758

15

0.238419

1.192093

0.244141

1.220703

0.305176

1.525879

16

0.119209

0.596046

0.122070

0.610352

0.152588

0.762939

Notes:

* For the top span, the 3 dB bandwidth is 1.15 times span shown.

To select sample frequency for time domain data, first divide the desired sample
frequency by 2.56 to convert it to a measurement span. Then locate the closest
measurement span on this table and choose the corresponding sample frequency at
the top of the table.

VT1433B User's Guide

Using the VT1433B

3-21

Advertising