8specifications, 1 operating environment, 2 other specifications – Olympus IPLEX UltraLite User Manual

Page 44

Advertising
background image

40

8

Specifications

8.1

Operating environment

CAUTION

• Meeting operating environment specifications does not constitute any guarantee against damage to or

malfunction of this equipment.

• Use AC adapter and battery charger indoors (normal temperature).

8.2

Other specifications

Operating temperatures

Insertion tube

In air

: -25 to 100

°

C

In water: 10 to 30

°

C

Other parts than above In air

: -10 to 40

°

C (Battery-powered operation)

0 to 40

°

C (AC adapter-powered operation)

Operating atmospheric pressure

Insertion tube

In air

: Normal pressure (1,013 hPa)

In water: 1,013 to 1,368 hPa

Other parts than above In air

: Normal pressure (1,013 hPa)

Operating environment humidity

All parts

15 to 90% (relative humidity)

Liquid resistance

Insertion tube

No trouble when exposed to machine oil, light oil or 5% saline solution.

Other parts than above

Waterproofing

Insertion tube

Waterproof construction. Can be used underwater with the optical adapter
attached.

Other parts than above Water resistant proof construction. Use in water is not supported. The instrument is

not water-resistant if the battery cover, SD card cover, or AC adapter connector cap
is open.

Optical system

For optical adapter specifications, see "8.3 Optical adapter specifications" (P. 43).

Distal end

Outer diameter

V84 Series:

φ4.0 mm

V86 Series:

φ6.0 mm

Distal end rigid section
length

See the distal end rigid section length specification of the optical adapter in use.

Angulation

IV8420U, IV8435U, IV8620U : UP, DOWN, RIGHT and LEFT each 130° *
IV8635U

: UP, DOWN, RIGHT and LEFT each 120° *

*when insertion tube is straight.

Flexible section

Flexibility

IV84 series: Fixed hard tube from the distal end to the main unit.
IV86 series: TF tube with flexibility gradually increasing toward the distal end.

Outer diameter

IV84 series:

φ4.0 mm

IV86 series:

φ6.0 mm

Index lines

An orange line is drawn 500 mm from the distal end. White lines are drawn at every
1,000 mm from the distal end. The number of white lines is increased at every
1,000 mm (1 line at 1,000 mm, 2 lines at 2,000 mm).

Total length

IV8420U: 2 m

IV8435U: 3.5 m

IV8620U: 2 m

IV8635U: 3.5 m

Advertising