Setup

49

Compensation Entry Methods

You can enter the segment boundaries and compensation factors using the

keypad, or you can have the readout calculate them for you. Normally, you
would use the keypad entry method only to restore or adjust a known
compensation value.

Manual Entry

This brings up the softkeys needed to let you enter compensation
and segment information using the keypad.

The

NEW

SEGMENT

softkey brings up the

DEFINE SEGMENT

form.

F

ACTOR

—Enter the compensation

factor in parts per million.

S

EGMENT

—You can change only the

TO

field. Use the

ENCODER

END

softkey if the

boundary is at the end of the
encoder. Use the

POINT

softkey to

enter the physical location of the
segment boundary. The current
reading head location may be entered using

TEACH

POSITION

softkey.

The

DONE

softkey saves the segment information in the table.

Continue adding segments with the

NEW

SEGMENT

softkey. You must add them in

order, from the encoder’s most negative end to its most positive end.

The

CLEAR

TABLE

softkey will delete the information from the table, erasing all

segments and compensation factors, so that the readout will not perform any
error compensation.

ENTER

COMP

FACTOR

PPM

2

DEFINE SEGMENT

SEGMENT

FROM

TO

ENCODER END

ENCODER END

NEW

SEGMENT

CLEAR TABLE

ANCHOR

SEGMENT 1

CANCEL

DONE