Position the standard in the center of the segment, Enter one edge of the standard, Enter the other edge of the standard – ACU-RITE VRO 300M User Manual

Page 58: Press

Advertising
background image

VRO 300M

Setup

51

You will need a dial indicator and a measurement standard.

1.

Position the standard in the center of the segment.

2.

Enter one edge of the standard.

3.

Enter the other edge of the standard.

4.

Enter the actual size of the standard, including the probe
diameter if necessary.

5.

Press

DONE

to save the segment information into the table.

The readout will automatically set the segment boundaries between the

current segment and the adjacent segments.

FIRST POINT

SECOND

POINT

8.1250

Opposite sides

FIRST POINT

SECOND

POINT

8.0000

Same side

If the calculated factor is 9999 or –9999, the machine
inaccuracy is too great and cannot be compensated.