System Configuration and Operation

Section 3

3.150

uS/cm



Km=1.124
25.0°C

11:33

AM

24.38°

C

1.50%/°C
T CELL

CELL

SETUP

DIAG

RANGE

OPERATION

Current cell constant

Offset

When measuring conductance below 10uS (resistance above 100K

Ω), the insulation resistance of

the cell leads may be a source of error. This error may be greatly reduced by allowing the 3200 to
calculate and store a zero offset value.

From the Calibrate K menu, press the [OFFSET] soft key to enter the Offset menu.

OFFSET

Offset value is only used below 10uS
0.0000uS offset
To measure, connect a clean, dry cell
and press MEASURE

REMOVE

MEASURE

Current offset


To measure and store the offset for the current cell in the currently selected configuration,
connect a clean, dry cell to the 3200, then, press the [MEASURE] soft key. The current offset is
displayed on the screen.

OFFSET

Offset value is used below 10uS
0.0015uS offset
To measure, connect a clean, dry cell
and press MEASURE

REMOVE

MEASURE

Current offset


To remove a stored offset (set the offset to zero), press the [REMOVE] soft key.

Send

From the Calibrate K menu, the [SEND] soft key may be used to send the current cell constant
data to a computer or serial printer via the RS232 port.

3.3 TEMPERATURE

COMPENSATION

There are four options for temperature compensation on the YSI 3200.

1. Directly enter a linear temperature compensation factor (0 - 10%/°C)

2. Select a standard non-linear compensation curve (ultrapure, natural).

YSI Incorporated

Model 3200

16