System Configuration and Operation

Section 3

1017

uS/cm

10.3

mS

OPERATION

984.6

Ωcm

97.077

K=0.098
25.0°C

691

mg/L
TDS

0.5

ppt
SAL

1.91%/°C
T CELL

CURRENT RANGE [95.0mS - 3.00S]

CELL

SETUP

DIAG

RANGE

conductance

resistance

conductivity

resistivity

If the [RANGE] soft key is pressed again, the hardware switches to the 0 - 0.9999uS range. The
conductance and resistance fields are now out of range and the 3200 displays a warning. The
conductivity, resistivity, TDS and salinity fields are still operating, but, since the hardware is out
of range, readings are no longer accurate.

924.5

uS/cm

0.9999

uS

OPERATION

1082

Ωcm

0.000

M

K=0.098
25.0°C

629

mg/L
TDS

0.4

ppt
SAL

1.91%/°C
T CELL

OUT OF RANGE

CELL

SETUP

DIAG

RANGE

out of range

warning


3.6 MAKING

MEASUREMENTS

After setting up the 3200 instrument and cell as described earlier, the following basic steps should
be used to make measurements.

1. Make sure that the 3200 is setup to read the units you wish to measure in (conductance,

conductivity, resistivity, etc.). To change the units see section 3.5 Display Configuration,
Measurement Units.

2. Verify that the 3200 is properly setup to use the current cell by measuring, or calibrating

with, standard conductivity solution(s). See section 3.2 Cell Calibration.

3. Immerse the cell (YSI 3200 series) in the solution to be measured. If using a YSI 3400

series cell and correcting for temperature, also place the temperature probe in the
solution.

4. Gently tap the cell to remove any air bubbles and dip the cell in the solution 2 or 3 times

to ensure proper wetting. The cell electrodes must be submerged and the electrode
chamber must not contain any trapped air. If using a flow through or fill cell, be certain it
is completely full.

5. Allow time for the temperature to stabilize.

6. Read the display, send the reading to a computer or serial printer, or store the reading in

memory.

7. Rinse the cell (and temperature probe, if used) with distilled or deionized water.

YSI Incorporated

Model 3200

25