System Configuration and Operation

Section 3

[0] # Configuration number

CONFIG


[CONFIG 0] Title

0.0000uS offset

[Single]

Cell

Cal

[1.00]

K

[Linear] Comp

[1.91] %/°C Comp val

[0.68] T.D.S. mult

[25.0] °C Comp Temp

ENTER TO SEND CONFIGURATION

UP

DOWN

DIGIT

NEXT

Press the [UP] or [DOWN] soft key to select one of the six (0-5) stored configurations. Each
configuration contains the single cell constant, zero offset, multiple cell constants, total dissolved
solids (TDS) multiplier, linear temperature compensation coefficient and non-linear temperature
compensation data.

To change the currently selected configuration, press [NEXT] to move the cursor to the field you
want to change. Then, use the [UP], [DOWN] and [DIGIT] soft keys to change the field, except
Offset. The Offset field can only be changed from the Offset menu. The following table shows
the choices for each field.

Example: To select the standard non-linear temperature compensation curve for ultrapure water,

press [NEXT] three times, to move the cursor to the Comp field, then, press the [UP]
or [DOWN] soft key to cycle through the selections until Pure is displayed. Press
[MODE] two times to exit to the Operation menu.

Field

Range / options

Description

Title

A - Z or 0 - 9

User defined name for this configuration.

Cell Cal

Single or Multi

Selects between a single fixed cell constant
and multiple cell constants.

Comp

Linear, N. Linear, Pure, Natural
or Off

Selects between a linear temperature
compensation coefficient, non-linear
temperature compensation curve, standard
non-linear temperature compensation curve
(pure or natural) or no temperature
compensation.

TDS mult

0.55 - 0.90

Factor used to calculate the total dissolved
solids from the conductivity.

uS offset

0.0000 - 0.0150

Zero offset. Used only when measuring
conductance below 10uS to offset cell lead
insulation resistance. This value can only be
changed from the Offset menu.

K

0.001 - 100.999

Single fixed cell constant value (K). This
value is only used when the Cell Cal field is
set to Single.

YSI Incorporated

Model 3200

13