Maintenance

Section

6

2. Use the [UP], [DOWN] and [DIGIT] soft keys to enter the platinization time required in

minutes. Six minutes is usually sufficient for most electrodes.

3. Immerse the cell in the platinizing solution (YSI 3140). Make sure that both electrodes

are submerged.

4. Press [ENTER] to start the platinization process. The following screen, showing the time

remaining and the polarity of the polarization current, will be displayed.

PLATINIZE


5.8 MIN remaining time

+ platinization polarity


PRESS MODE TO ABORT PLATINIZATION

5. After the platinization process is complete, remove the cell from the platinizing solution.

If you want to stop the platinization before the timer reaches zero, press the [MODE] key
to abort.

6. Thoroughly rinse the cell with distilled or deionized water.

7. Promptly return the platinizing solution to its container.