Warranty and Repair

Section 9

Temperature Verification

Temperature

Cell Connector Pin 1 & 2
Input Resistance **

EX. TEMP Connector
Input Resistance **

Displayed
Temperature

0

o

C

32660

Ω ± .25%

19600

Ω ± .25%

0.00 ± .15

o

C

25

o

C

10000

Ω ± .25%

6000

Ω ± .25%

25.00 ± .15

o

C

40

o

C

5329

Ω ± .25%

3197

Ω ± .25%

40.00 ± .15

o

C

70

o

C

1752

Ω ± .25%

1051

Ω ± .25%

70.00 ± .20

o

C

100

o

C

678.5

Ω ± .25%

407.1

Ω ± .25%

100.00 ± .20

o

C

** Input temperature resistances either via the EXT. TEMP connector or via pins 1 & 2 of the