Temperature Verification
Temperature
Cell Connector Pin 1 & 2
Input Resistance **
EX. TEMP Connector
Input Resistance **
Displayed
Temperature
0
o
C
32660
Ω ± .25%
19600
Ω ± .25%
0.00 ± .15
o
C
25
o
C
10000
Ω ± .25%
6000
Ω ± .25%
25.00 ± .15
o
C
40
o
C
5329
Ω ± .25%
3197
Ω ± .25%
40.00 ± .15
o
C
70
o
C
1752
Ω ± .25%
1051
Ω ± .25%
70.00 ± .20
o
C
100
o
C
678.5
Ω ± .25%
407.1
Ω ± .25%
100.00 ± .20
o
C
** Input temperature resistances either via the EXT. TEMP connector or via pins 1 & 2 of the
mini DIN connector, not both. See the following figure, Test and Verification.