Data Storage and Retrieval

Section 4

The baud rate range is 1200 to 19200. After you have selected the parameters, press [MODE]
once, to return to the Send Setup menu, or three times to return to Operation mode.

To select the report format and which data fields will be sent over the RS232 serial port when the
[SEND] soft key is pressed, press the [SET SEN] soft key. The following screen will be
displayed.

[ ] Field delimiter

SETUP SEND

[ ] Space units

[*] Temperature

[*]

Date

[Conductance]

[*]

Time

[Conductivity]

[ ] Cell config

[ ] Compensation

[0] S interval

UP

DOWN

DIGIT

NEXT

Each parameter is selected by placing an asterisk character between the brackets in front of it.
Press the [NEXT] soft key to select the parameter that you want to change. Then press the [UP] or
[DOWN] soft key to turn the asterisk character on or off.

You can select the report format that you wish to use by changing the field delimiter. The default
is a space [ ]. This provides a standard text output. You may also select a comma [,] by pressing
the [UP] or [DOWN] soft key.

When Space units is selected, by placing an asterisk character between the brackets in front of it
[*], a space will be inserted between the value and the units. This is useful when importing the
data to a spreadsheet.

The following are examples of each send format:

Space Delimited Format, send (current display)

1/23/97 12:34:56AM 0# 1.23%/°C 12.34C 12.3mS 1.23ppt

12/12/97 12:34:56AM 0# 1.23%/C 12.34C 1234uS/cm 0.12ppt

Comma Delimited Format, send (current display)

1/23/97, 12:34:56AM, 0# ,1.23%/°C, 12.34C , 12.3mS ,1.23ppt

12/12/97, 12:34:56AM, 0# ,1.23%/C, 12.34C , 1234uS/cm ,0.12ppt

Space Delimited Format, send (recalled data)

0 3/20/97 5:37:17PM 1234uS/cm 23.45C K=0.09 1.91%/C 9999uS

1 3/20/97 5:46:15PM 1234uS/cm 23.45C K=0.09 1.91%/C 9999uS

2 3/20/97 5:57:34PM 1234uS/cm 23.45C K=0.09 1.91%/C 9999uS