Beeper – YSI 3200 User Manual

Page 26

Advertising
background image

System Configuration and Operation

Section 3

1. From Operation mode, press the [SETUP] soft key, then, press [SYSTEM] to enter the Setup
Sys menu. The following menu will be displayed.

[Yes] key beep

SETUP SYS


[Off] alarm status
[0.00] low

[Off] beep on alarm

[0.00] high

[On] send on alarm

[Conductance] uS alarm source

UP

DOWN

DIGIT

NEXT

2. Use the [NEXT] soft key to select the Alarm status field, then press [UP] or [DOWN] to

change the Alarm status to “On”.

3. Press [NEXT] to select the Low field, then use the [UP], [DOWN] and [DIGIT] soft keys to

enter the set point for the low alarm.

4. Press [NEXT] to select the High field and use the [UP], [DOWN] and [DIGIT] soft keys to

enter the set point for the high alarm.

5. Press [NEXT] to select the alarm source, then press [UP] or [DOWN] to change the source.

The source may be conductance, conductivity or resistance.

6. Press [NEXT] to select “beep on alarm”, then press [UP] or [DOWN] to turn the alarm beeper

on or off.

7. Press [NEXT] to select “send on alarm”. When this parameter is enabled (On), readings will

be sent to a computer or serial printer whenever the alarm is triggered.

Press [MODE] once, to return to the Setup menu, or two times to return to Operation mode.

NOTE: To setup the communications parameters or select the report format and which data fields
will be sent over the RS232 serial port, see 4.3 Send, Send Setup.

BEEPER

The 3200 contains a beeper to indicate when a key has been pressed. The beeper may be turned
on or off from the Setup Sys menu. Press the [SETUP] soft key, while Operation mode, then
press [SYSTEM] to enter the Setup Sys menu .

Beeper status

[No] key beep

SETUP SYS


[Off] alarm status
[0.00] alarm low

[Off] beep on alarm

[0.00] alarm high

[On] send on alarm

[Conductance] uS alarm source

UP

DOWN

DIGIT

NEXT

YSI Incorporated

Model 3200

22